Electrostatic discharge (ESD) IEC 61000-4-2 | ± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air | ± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air | Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30% |
Electrical fast transient/burst IEC 61000-4-4 | ± 2 kV for power supply lines ± 1 kV for input/ output lines 100 kHz repetition frequency | ± 2 kV for power supply lines ± 1 kV for input/ output lines 100 kHz repetition frequency | Mains power quality should be that of a typical commercial or hospital environment |
Surge IEC61000-4-5 | Line to line: ± 0.5 kV, ± 1 kV Line to ground: ± 0.5 kV, ± 1 kV, ± 2 kV | Line to line: ± 0.5 kV, ± 1 kV Line to ground: ± 0.5 kV, ± 1 kV, ± 2 kV | Mains power quality should be that of a typical commercial or hospital environment |
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 | 0% residual voltage for 0.5 period at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0% residual voltage for 1 period at 0° 70% residual voltage for 25 periods at 0° Voltage interruptions: 0% residual voltage for 250 periods at 0° | 0% residual voltage for 0.5 period at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0% residual voltage for 1 period at 0° 70% residual voltage for 25 periods at 0° Voltage interruptions: 0% residual voltage for 250 periods at 0° | Mains power quality should be that of a typical commercial or hospital environment. If the user of the Nio Color 5MP requires continued operation during power mains interruptions, it is recommended that the Nio Color 5MP be powered from an uninterruptible power supply or a battery |