4.17.4 DICOM calibration

About DICOM calibration

The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve as well as possible.

To start DICOM calibration:
  1. Bring up the OSD main menu.
  2. Navigate to the Configuration > Calibration > Embedded QA menu.
  3. Select DICOM calibration to start the calibration.

    Warning: Pressing a key during calibration will abort the calibration, previous values will be restored.
    Note: After calibration, the compliance check will start automatically.