E.2 Laser pulse calibration

Calibration introduction

If the projector is to be used only on one laser power setting, use this guide for calibration:

  1. Allow unit to run for at least 15 minutes to enable stable light output.
  2. Select RealColor mode preset: Native. Menu -> Image -> Advanced settings -> Realcolor P7

    Image E–1
  3. Select desired BrilliantColor mode. Menu -> Image -> Advanced settings -> BrilliantColor

    Image E–2
  4. Select desired Smear reduction mode Menu -> Image -> Advanced settings -> Smear reduction

    Image E–3
  5. Use a white test image either from an external source or the internal White test pattern. A chromameter is needed to measure the coordinates of the white test pattern.
  6. Adjust the laser pulse value to obtain the desired white y-coordinate.

    1. Log in as Power user. Go to menu Menu -> Settings -> Maintenance -> Advanced

      Image E–4
    2. Enter the access code from the remote controller numeric pad to have access to the feature.

      Image E–5
    3. Enter the menu Maintenance -> Advanced -> Laser pulsing

      Image E–6
      Image E–7
  7. Go to Native RealColor within the Maintenance section to input the resulting white x and y coordinates. Click Apply when done.

    Image E–8